Navigation
Cart£0.00

Miniature Zirconium Dioxide Oxygen Sensors

Return to Previous Page
Specification

TECHNICAL SPECIFICATIONS

Heater voltage
    Standard response sensor4Vdc +/- 0.1Vdc (1.7A)
    Standby1.65Vdc (0.7A)
    Fast response sensor4.35Vdc +/- 0.1Vdc (1.85A)
    Standby2Vdc (0.85A)
Pump impedance at 700°C< 6kΩ
Permissible gas temperature-100 to +250°C
Gas flow rate0-10 m/s
Repetitive permissible acceleration5g
Incidental permissible acceleration30g

OUTPUT VALUES

Oxygen pressure range2 mbar-3bar max.
Accuracy5 mbar max.
Internal operational temp700°C
Response time (10-90% step)
    Standard response sensor < 15s
    Fast response sensor < 4s
Warm up time (prior to sensor operation)60s
Warm up time (from standby)20s
Output stabilisation time~ 180s


OUTLINE DRAWINGS

Mini_diagrams

ELECTRICAL INTERFACE

Mini_elec-connect

Pin Designation
P Pump
C Common
H1 Heater (1)
S Sense
H2 Heater (2)

Features & Benefits

FEATURES

  • Zirconium dioxide technology
  • Requires remote electronics to operate
  • Integral sensing element
  • Small in size 12mm diameter
  • Stainless steel construction
  • Standard (15s) or fast (4s)response time options available
  • High accuracy
  • Operating oxygen range from 0.1% to 100%

BENEFITS

  • Long life, non-depleting technology
  • No reference gas required creating the ability to measure a wide oxygen range
  • No need for temperature stabilisation
  • Linear output signal
  • Can be used in high pressure and temperature environments
  • Simple recalibration if/when required
  • Full technical and application support available

Downloads Sample/Quote Request

Sample/quote request

  • This field is for validation purposes and should be left unchanged.

Related Products

temp-footer-ctas